User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: TA11 Frequency Response Sub DATE: 16-Jun-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 66 CONFIGURATION: HP 3325A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 MATH mem1 = m[1] 1.002 IEEE #[MEM1]:P=125;[D250] 1.003 DISP Connect the 3325A SIGNAL output to Channel [MEM1] 1.003 DISP on the UUT. 1.004 IEEE [@3325] FU 1; FR 10 HZ; AM 2.5 VO [GTL] 1.005 IEEE SP; 1.006 DISP Using the MODIFY keys on the 3325A, adjust the amplitude 1.006 DISP from the generator so the sigal completely fills the 1.006 DISP UUT print out without clipping. 1.007 IEEE HLT; 1.008 IEEE [@3325]IAM[I] 1.009 MEME >20 1.010 IEEE [@3325] FU 1; FR 10 KH; AM [MEM] VO [GTL] 1.011 IEEE SP; 1.012 IEEE [D5000]HLT; 1.013 DISP Count the number of full, and/or, fractional part, of 1.013 DISP major divisions the signal is away from the left edge 1.013 DISP of the UUT print out. 1.014 MEMI Enter the number of divisions counted. 1.015 MATH mem = m[20] - (mem * .1) 1.016 MEME 1.017 MEME <20 1.018 MEMC V 10.00% 10kH 2.001 MEME <20 2.002 IEEE [@3325] FU 1; FR 25 KH; AM [MEM] VO [GTL] 2.003 IEEE SP; 2.004 IEEE [D5000]HLT; 2.005 DISP Count the number of full, and/or, fractional part, of 2.005 DISP major divisions the signal is away from the left edge 2.005 DISP of the UUT print out. 2.006 MEMI Enter the number of divisions counted. 2.007 MATH mem = m[20] - (mem * .1) 2.008 MEME 2.009 MEME <20 2.010 MEMC V 30.00% 25kH 3.001 MEME <20 3.002 IEEE [@3325] FU 1; FR 5 HZ; AM [MEM] VO [GTL] 3.003 MATH mem1 = m[1] 3.004 IEEE #[MEM1]:FIL=Y;[D250] 3.005 IEEE SP; 3.006 IEEE [D5000]HLT; 3.007 DISP Count the number of full, and/or, fractional part, of 3.007 DISP major divisions the signal is away from the left edge 3.007 DISP of the UUT print out. 3.008 MEMI Enter the number of divisions counted. 3.009 MATH mem = m[20] - (mem * .1) 3.010 MEME 3.011 MEME <20 3.012 MEMC V 20.00% filt=on 4.001 IEEE [@3325] FU 1; FR 60 HZ; AM [MEM] VO [GTL] 4.002 IEEE SP; 4.003 IEEE [D5000]HLT; 4.004 EVAL Is the {trace less than 1 mm pk-pk at 60 Hz w/the filter on}? 5.001 END